SERVOFLEX MiniMP (5200 Multipurpose) | Portable Gas Analyzer

SERVOFLEX MiniMP (5200 Multipurpose) | Portable Gas Analyzer

Date Submitted: 24-02-2021 03:02:27

 5200
PORTABLE BENCHTOP GAS ANALYSER     
                            
For analysing the composition of common gas mixtures in laboratories and light industrial applications    
  • Non-depleting paramagnetic and infrared measurement technology ensures the unit is always ready for use
  • Accurate, reliable and component specific, giving confidence that observed values are credible and not due to background interference
  • Customer pre-tested software offers ease of use, simple calibration and advanced features
  • Innovative construction gives a robust unit with a space saving footprint
  • Rechargeable battery and vehicle power (12/24V) options enable mobile use
  • TÜV and MCERTS approved to EN15267-3 (certification of automated measuring systems) available

SPECIFICATION

 

Gases measured
Oxygen (O2) and/or Carbon Dioxide (CO2) 
SENSORS OXYGEN Carbon Dioxide 
Technology: Paramagnetic Infrare
Variant: Standard Industrial High accuracy All variants
Full scale range 0-100% O2 0-100% O2 0-100% O2 10%, 25%, 50%, 100%
Minimum output range: 0-10% O2 0-1% O2 0-1% O2 0-10% FSR
Cell construction: Polymer (PPS) 316 Stainless Steel 316 Stainless Steel 316 Stainless Steel
Decimal places:

1

1

2

Ranges:<10%:2 >10%:1
PERFORMANCE 
Accuracy, general use: ±0.2% O2 ±0.1% O2 ±0.05% O2 ±2% FSR
Accuracy,stable environment: ±0.1% O2 ±0.1% O2 ±0.02% O2 ±1% FSR
Zero drift, per week: ±0.4% O2 ±0.2% O2 ±0.2% O2 ±4% FSR
Response time (T90): <10s <15 seconds <15 seconds <10 seconds
Tilt effect, at 15° from cal: ±0.3% O2 ±0.15% O2 ±0.15% O2 ±1% FSR
Power cycle offset: ±0.4% maximum N/A N/A N/A
SAMPLE
Sample gas: Clean, dry, non-flammable and non-toxic gases only.
Note: Though samples containing >5% CO2 are toxic they can be analysed if suitable precautions are taken.
Flow control: To maximise measurement stability, unpumped units are supplied with an automatic flow control device (AFCD)
Over the specified inlet pressure range this controls sample flow rate to approximately 1.5 to 6 litres (0.05 to 0.2 cubic feet) per minute
 